Survey and analysis of the results of creatinine measurement by 31 detection systems of 13 hospitals in Beijing
- VernacularTitle:北京市13家医院31套血清肌酐检测系统测定结果的比对调查及分析
- Author:
Jing XU
;
Guo-Bin XU
;
Qing TONG
;
Hui-Min JIA
;
Qing-Tao WANG
;
- Publication Type:Journal Article
- Keywords:
Creatinine;
Data collection;
Laboratories
- From:
Chinese Journal of Laboratory Medicine
2003;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective Understand the variation of serum creatinine measurement in clinical laboratories of some hospitals in Beijing.Methods 8 samples of mixed frozen human serum added different creatinine concentration standard materials(the creatinine concentration were80-1 000 ?mol/L)and 8 samples of mixed frozen serum of patients(contained different creatinine concentration)were distributed to 13 clinical laboratories(31 series of detection systems)with the way of spot investigation.Every clinical laboratories measured the samples followed the standard operating procedure.Results As to the mixed frozen human serum added different creatinine standard materials,the CV of different detection systems results were 5.74%-9.68%;as to the mixed frozen patients' serum,the CV was 5.90%-11.69%. Compared with Beckman closed detection systems,the results of Dade systems(which used the kinetic alkaline pieric acid method)showed the bias were-5.99%-0.35%,and as to the other systems which measured by alkaline picric acid method,when creatinine concentrations were 200 ?mol/L,the results showed negative bias,and the greatest bias was-8.45%.The bias plots revealed negative for all of the detection systems with enzymatic method over the whole concentration range,and the greatest bias was -8.88%.Conclusions The creatinine determination results of Beckman and Dade closed detection systems were consistent.The results of detection systems which used enzymatic method were generally lower than Beckman detection systems.What's more,the creatinine measurement variations of clinical laboratories were very large,especially for the results of unclosed detection systems,so it was urgent need to solve the standardization of creatinine measurement.