Comparison of routine test kits of creatinine in human serum with reversed-phase high-pressure liquid chromatographic assay
- VernacularTitle:几种人血清肌酐常规测定方法与反相高压液相色谱测定法的比对研究
- Author:
Jing XU
;
Qing-Tao WANG
;
Guo-Bin XU
;
- Publication Type:Journal Article
- Keywords:
Chromatography,high pressure liquid;
Creatinine;
Intermethod comparison
- From:
Chinese Journal of Laboratory Medicine
2000;0(06):-
- CountryChina
- Language:Chinese
-
Abstract:
98.95%.The results of certified reference materials were consistent with the target value,and average deviation was -0.31%~1.35%.HPLC was served as the independent variable.When ereatinine was 200 ?mol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -10%~13%,-13%~14% and -20%~10%,respectively.Bias of enzymatic method results was mostly negative,when creatinine was