The study of detecting and correcting part of biochemical markers by"B&G System"
- VernacularTitle:用“B&G System”检测与校正部分生化指标研究
- Author:
Min QU
;
Xi-Xin WU
;
De-Hui LIU
;
- Publication Type:Journal Article
- Keywords:
Glomerular filtration rate;
System integration;
Biochemical marker
- From:
Chinese Journal of Primary Medicine and Pharmacy
2005;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To verify the accuracy of detecting part of biochemical markers of the new"B&G System".Methods Blood serum creatinine(SCr),urea nitrogen(BUN),serum uric acid(UA),Ca or P of the same serum example were detected by"B&G System"and Hitachi 7170S automatic biochemistry analysis instrument (7170S) respectively.Each biochemical marker was divided into two teams on the basis of measure values of 7170S, and then measure values of two methods were compared.If there were significant differences,the biochemical mark- ers's detection program of"B&G System"was corrected,and then measure values of two methods of corresponding biochemical markers were compared.Results Two methods' measure values of SCr of abnormal team,BUN of each team,SUA of each team and P of each team weren't significantly different,measure values of SCr of normal team,Ca of two teams were significantly different.After corrected"B&G System",two methods' measure values of sCr of normal team,Ca of two teams weren't significantly different.Conclusion"B&G System"through correction could accurately detect biochemical markers,and it's worth applicating in clinic.