Use of the SmartPrep technique with concentration of contrast medium control in CT angiography of vertebral and carotid artery
- VernacularTitle:对比剂密度监控智能扫描技术在椎动脉及颈动脉CT血管成像中的应用
- Author:
Kang-Wei SHAO
;
Cheng-You GONG
;
Wei LIU
;
Li-Xin YUAN
;
- Publication Type:Journal Article
- Keywords:
Vertebral artery;
Carotid artery;
Tomography,X-ray computed;
Angiography
- From:
Chinese Journal of Radiology
2001;0(07):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the optimal enhancement phase and exact triggering scan time on spiral CT angiography of vertebral and carotid artery,and to improve the rate of 3D-CT display in vertebral and carotid artery.Methods One hundred and thirty-three cases of CT angiography of vertebral and carotid artery(87 vertebral and 46 carotid artery)were examined with SmartPrep software,which controled the scan at real time after 100 ml contrast medium(350 mg I/ml Omnipaque)were quickly injected through foot back vein.One hundred and thirteen cases of control group were examined by triggeing off the scanning with a delay time of 18—22 s after blous injection through foot back vein.The vertebral and carotid arteries were demonstrated with 3 dimensional reconstruction.Results The successful rate in SmartPrep group was 100% with the enhancement situated in best(n=130)or good(n=3)scanning phase.The threshold concentration values of target vessel were kept on 100 HU in the whole scan period[(18?4)s].However, the triggering time of enhancement scanning was not equal(15—43s)and the deviation was up to 28 s.The average concentration of target vessel was obviously higher in SmartPrep group[(161.4?2.0)HU]than in the control group[(133.3?2.2)HU](t=-9.456,P