Evaluation of predictive performance of Diprifusor target-controlled infusion system for propofol in Chinese patients
- VernacularTitle:国人应用异丙酚靶控输注系统准确性的评价
- Author:
Yu-Hong LI
;
Jian-Jun YANG
;
Jie TIAN
;
Al ET
- Publication Type:Journal Article
- Keywords:
Propofol;
Drug delivery systems;
Pharmacokinetics
- From:
Chinese Journal of Anesthesiology
1995;0(12):-
- CountryChina
- Language:Chinese
-
Abstract:
25% or decrease by 15%,at termination ofTCI propofol and 5 rain after termination for determination of arterial plasma propofol concentrations which werecompared with the values predicted by the TCI system.Performance was determined by the median performanceerror(MDPE),the mean absolute performance error(MDAPE),the divergence and the wobble.Results A totalof 200 blood samples were taken.The mean value for MDPE was 14.9%(-21.6%-42.9%),for MDAPE23.3%(6.9%-62.5%),for divergence -1.9%?h~(-1)(-32.7%?h~(-1)-23.0%?h~(-1))and for wobble 18.9%(4.2%-59.6%).PE did not increase with time but with increasing target propofol concentration.ConclusionThe anesthesia was satisfactory in all patients.The predictive performance of Diprifusor TCI system is consideredacceptable for clinical purposes.The wobble is relatively big,for Chinese patients the pharmacokinetic model needsimproving.