Detection and analysis of myeloperoxidase level in different types of samples
10.3969/j.issn.1673-4130.2016.23.003
- VernacularTitle:不同标本类型间髓过氧化物酶水平的检测分析
- Author:
Yongbo WANG
;
Jing QIAN
;
Huilin CHEN
- Keywords:
myeloperoxidase;
anticoagulant;
types of samples;
reference intervals
- From:
International Journal of Laboratory Medicine
2016;37(23):3246-3248
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the difference of myeloperoxidase(MPO)detection results among different types of samples , anticoagulant selection and comparison of detection results .Methods The plasma samples containing EDTA‐K2 or heparin and ser‐um samples of common biochemical tube without containing anticoagulant were collected from 165 persons undergoing healthy physical examination .The MPO level in 3 kinds of sample was detected .Then the detection results in each group were performed the statistical analysis .Results The MPO detection results had statistical difference between plasma samples with different antico‐agulants and between plasma samples with different anticoagulants and serum sample without anticoagulant in the same person (P<0 .05) .Conclusion Different samples types have larger difference in M PO detection results .It is suggested that each laboratory should formulate the corresponding reference intervals according to different anticoagulants ;because EDTA‐K2 anticoagulation plasma is not affected by the release of MPO from in vitro leukocytes ,it is recommended to adopt EDTA‐K2 anticoagulation blood as the first selected of M PO level detection .