The difference of C reactive protein level detected by immunity transmission turbidity and dry immunofluorescence
10.3969/j.issn.1673-4130.2015.16.049
- VernacularTitle:免疫透射比浊法与干式免疫荧光法检测C反应蛋白的结果差异
- Author:
Lijing WANG
- Publication Type:Journal Article
- Keywords:
immunity transmission turbidity;
dry immunofluorescence;
C reactive protein
- From:
International Journal of Laboratory Medicine
2015;(16):2406-2407
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the difference of C reactive protein level detected by immunity transmission turbidity and dry immunofluorescence.Methods The clinical data of 56 patients with various kinds of infectious diseases who were treated in our hospital from September 2012 to January 2014 was retrospectively analyzed.Results The levels of C reactive protein in the obser-vation group detected by immunity transmission turbidity and dry immunofluorescence were significantly lower than those in the control group (P <0.05),and the positive rate of C reactive protein was significantly higher than that in the control group(P <0.05).Conclusion The levels of C reaction protein detected by immunity transmission turbidity and dry immunofluorescence in pa-tients with a variety of acute infectious diseases show no significant difference,suggesting that immunity transmission turbidity can be used as an alternative to dry immunofluorescence.