Determination of Fe, Co, Mn and Ni in Synthetic Diamonds by Inductively Coupled Plasma Atomic Emission Spectrometry
10.3724/SP.J.1096.2014.31021
- VernacularTitle:电感耦合等离子体原子发射光谱法测定人造金刚石中铁、钴、锰、镍
- Author:
Guang-Jie-Zi GAO
;
Yan-Ping LI
;
Sheng-Ya FENG
;
Yan-Jun XIE
;
Chun-Jiang KUANG
;
Cheng CAO
;
- Publication Type:Journal Article
- Keywords:
Inductively coupled plasma atomic emission spectrometry;
Synthetic diamond;
Iron;
Cobalt;
Manganese;
Nickel
- From:
Chinese Journal of Analytical Chemistry
2014;(3):457-458
- CountryChina
- Language:Chinese
-
Abstract:
A method for the determination of Fe, Co, Mn and Ni in synthetic diamonds by inductively coupled plasma atomic emission spectrometry ( ICP-AES) was proposed. The synthetic diamond sample was decomposed completely, while the sample was burned in air at 1000 ℃ for 10 h, and then a mixed acid of H2 SO4 , aqua regia and HClO4 was used for the dissolving the residue of the sample. In this method, the limits of detection of Fe, Co, Mn and Ni were 0. 0147, 0. 0018, 0. 0006 and 0. 0027 mg/L, respectively. Under the optimum condition, Fe, Co, Mn and Ni in synthetic diamond sample were determined. The values of RSDs (n=7) were less than 0. 5%. The recoveries of added standard were 94. 0%-105. 0%.