Comparison Research between Fiber-optic Method and Imported Drug Registration Standard Method for Dissolution of Doxazosin Mesylate Extended Release Tablets
- VernacularTitle:光纤法与进口药品注册标准法测定甲磺酸多沙唑嗪缓释片释放度的比较
- Author:
Shanshan GONG
;
Jin GAO
- Publication Type:Journal Article
- Keywords:
Fiber-optic method;
Imported drug registration standard method;
Doxazosin mesylate extended release tablets;
Dis-solution
- From:
China Pharmacist
2015;18(10):1682-1684
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To compare the dissolution results of doxazosin mesylate extended release tablets determined by a fiber-optic method and the imported drug registration standard method. Methods:The drug release process was determined directly with a FODT-601 fiber-optic medicine dissolution/ release rate process monitoring system. Aqueous solution of hydrochloric acid and sodium chloride was used as the dissolution medium, the paddle rotation rate was 75 r·min-1 , the detection wavelength was set at 246 nm while the reference wavelength was 550 nm. The detection length was 5 mm. Results:The standard curve of doxazosin mesylate was linear within the concentration range of 0. 468 1-11. 700 0μg·ml-1 and r values were greater than 0. 999 0. The intra-and inter-day RSD (n=6) was 1. 6% and 2. 0%, respectively. The recovery of doxazosin mesylate was 99. 0% and RSD was 1. 4(n=9). The results determined by the fiber-optic method were generally higher than those determined by the standard method, and there were some differences in the measurement results. Conclusion:Fiber-optic method shows the whole dissolution process objectively. It is particularly prominent in the research on the dissolution of rapid, extended and controlled release preparations. However, it can not replace the standard method yet.