Ultrasonic quantitative analysis of the flexion cornal hip in the normal infant
10.3760/cma.j.issn.1004-4477.2017.08.014
- VernacularTitle:正常婴儿屈髋冠状切面的超声定量分析及其临床意义
- Author:
Lili MIAO
1
;
Qinghua LIU
;
Xiaofang LIU
;
Xincun ZHANG
;
Yuxi YAN
;
Huanping PANG
Author Information
1. 山东大学齐鲁儿童医院超声科
- Keywords:
Ultrasonography;
Infant;
Hip joint;
Position
- From:
Chinese Journal of Ultrasonography
2017;26(8):703-707
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the influencing factors and clinical significance of ultrasonic quantitative analysis of the hip-flexion coronal section in the normal infants.Methods Totally 100 normal infants were enrolled.Angle α,angle β and FHC on different positions(mild-flexion and flexion of the hip) were mearured.And angle β on different points(the labrum central and the acetabular tips) were measured.The variation of the measurement index between different positions were analyzed.Results ①There was no significant difference in angle α between the neutral position and hip-flexion position (P >0.05).② FHC decreased in the hip-flexion position,and there was significant difference compared with the neutral position (P <0.05).③βc were greater than βt in the two positions (P <0.05);βc and βt were all greater in hip-flexion position than those in neutral position,the difference was statistically significant (P <0.05).Conclusions The results obtained from angle α is stable under the coronal flexion view of the normal infants,and does not vary with changing position.The change of the angle β and FHC with the hip flexion could be used to evaluate the stability of the hip.Measured angle β on the labrum tip has good repeatability.So this point should be selected to measure the angle β.