Analysis for comparative results of isotope dilution mass spectrometry in CDC-CRMLN project
10.13602/j.cnki.jcls.2017.09.13
- VernacularTitle:同位素稀释质谱法在美国CDC-CRMLN计划中的比对结果分析
- Author:
yan Wei ZHOU
1
;
xia Hong LI
;
bao Chuan ZHANG
;
yue Rui YANG
;
tao Jiang ZHANG
;
jiao Tian ZHANG
;
Ying YAN
;
Jie ZENG
;
Shu WANG
;
Jun WEN-XIANG
;
DONG CHEN
Author Information
1. 北京医院国家老年医学中心卫生部临床检验中心
- Keywords:
Centers for Disease Control and Prevention;
Cholesterol Reference Method Laboratory Network;
cholesterol;
triglycerides;
isotope dilution liquid chromatography tandem mass spectrometry
- From:
Chinese Journal of Clinical Laboratory Science
2017;35(9):690-695
- CountryChina
- Language:Chinese
-
Abstract:
Objective To summarize and review the comparative results of the isotope dilution mass spectrometry in Cholesterol Reference Method Laboratory Network (CRMLN)project of the Centers for Disease Control and Prevention (CDC)of the United States in order to provide quality controls for determination of blood lipid.Methods The isotope dilution liquid chromatography tandem mass spectrometry (ID-LC/MS/MS)methods for determination of serum total cholesterol (TC)and triglycerides (TG)were developed to participate in the comparison of CRMLN.The survey was conducted every three months before 2016 and every half year from 2016.Four kinds of reference materials with two parallel tubes for each material and each tube in duplicate were determined in every survey.At least two certified reference materials used as quality control samples were analyzed simultaneously in each determination.Results In the 15 comparisons the CV of the TC determination method in our laboratory was 0.43% while the CV of all the participated laboratories in CRMLN was 0.42%.The bias from the overall mean value in our laboratory was 0.22% while the bias from the CDC target values was 0.58%.The CV of the TG determination method in 15 tests of our laboratory was 0.62% and the bias was-0.98% from the overall mean value and-0.80% from the target values of CDC.Among the 60 results for comparison,98% (59/60)of CV in the TC determination met with CDC requirement for precision (CV≤ 1%),and 70% of bias (42/60) of the results met with CDC requirement for accuracy (bias ≤ 1%).For the 60 results in comparison of TG determination,92% (55/60)of bias of the results met with the accuracy requirement of CDC (bias ≤2.55%).Conclusion In CRMLN comparison the results of TC and TG determined by ID-LC/MS/MS method were consistent with the values which were certified by CDC and determined by other network laboratories.The comparative surveys may play an important role in the standardization of lipid determination,and should be expected to provide experiences and technical supports in the comparative plan for reference measurement in our country.