Evaluation of the targeted uncertainty in routine coagulation assay using EQA data
10.3760/cma.j.issn.1009-9158.2017.12.008
- VernacularTitle:采用室间质评数据建立常规凝血项目的目标测量不确定度
- Author:
Wei WANG
1
;
Yanyan QIN
;
Hongyi YIN
;
Qing TONG
;
Yufang LIANG
;
Yuhong YUE
;
Qingtao WANG
;
Rui ZHOU
Author Information
1. 100015,首都医科大学附属北京地坛医院输血科
- Keywords:
Diagnostic tests,routine;
Blood coagulation tests;
Reference values;
Uncertainty;
Laboratories,hospital;
Laboratory proficiency testing
- From:
Chinese Journal of Laboratory Medicine
2017;40(12):942-946
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish the target measurement uncertainty(MU)of the routine coagulation assay according to the External Quality Assessment data(EQA)of routine coagulation assay. Methods Beijing Center for Clinical Laboratory(BCCL)established the target measurementuncertainty for routine coagulation assayswith the"up-down"methodon the basis of 93 clinical laboratoriesEQA datain BeijingThese assays includedActivated partial thromboplastine time(APTT), Fibrinogen(FBG), International Normalized Ratio(INR), Prothrombin time(PT), Thrombin time(TT)and D-dimer, Compared with CLIA′88,the proficiency of current coagulation assayswas observed.Results The MU of six routine coagulation assayscompared with CLIA ′88 showed that: The 90th percentile MU met the creteriain APTTof group B,FBG of group A&B&C,INR of group B and D-dimer of group B.The 75th percentile MU met the creteriainINR of group A&C,PT of group C.The medium met the creteriainAPTT of group A&C,PT of group A and INR of group D.Conclusions Target Measurement Uncertainty was establishedin routine coagulation assay by using EQA data only,whichcan simplify the procedure of determining MU and continuously update MU according to the frequency of EQA.It has good clinical practical value.However, the applicability of this method should also be considered.