Clinical application value of transesophageal echocardiography and LAA CTA before percutaneous left atrial appendage closure with the Watchman occluder
10.3760/cma.j.issn.1004-4477.2017.11.011
- VernacularTitle:经食管超声心动图及CT左心耳造影在左心耳W atch m an封堵术前的应用价值
- Author:
Tao ZHANG
1
;
Libin CHEN
;
Huimin CHU
;
Feng MAO
;
Fei YU
;
Xueli ZHU
Author Information
1. 315010,浙江省宁波市第九医院超声科
- Keywords:
Echocardiography,transesophageal;
Left atrial appendage occlusion;
Computed tomography angiography;
Watchman occluder
- From:
Chinese Journal of Ultrasonography
2017;26(11):964-969
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the clinical application value of 2-dimentional transesophogeal echocardiography (2D-TEE) ,real-time 3-dimensional transesophogeal echocardiography(RT-3D TEE) and left atrial appendage CT angiography ( LAA CTA ) before percutaneous left atrial appendage closure ( LAAC) with the Watchman occluder . Methods Consecutive 50 atrial fibrillation ( AF) patients underwent 2D-TEE ,RT-3D TEE and LAA CTA examination before LAAC . The number of LAA lobes ,the diameter of landing zone (DLZ) and the depth of LAA were measured by using different methods . The correlation between LAA landing zone diameters measured by different methods and device size were calculated . Results Among 50 patients ,there was no significant difference in lobe nubmers between RT-3D TEE and LAA CTA ( P >0 .05) . The maximum diameter of landing zone(DLZ) from LAA CTA was greater than those from RT-3D TEE and 2D-TEE ( t =3 .977 ,5 .373 ,both P<0 .05) ,and DLZ from RT-3D TEE was greater than that from 2D-TEE ( t=2 .124 , P <0 .05) . The value measured by RT-3D TEE was greater than that by 2D-TEE for the LAA minimum DLZ ( t =2 .142 , P < 0 .05) . LAA depth:CTA value was greater than 2D-TEE and RT-3D TEE values( t =2 .674 ,4 .066 ,both P < 0 .05) ,and 2D-TEE value was greater than RT-3D TEE value( t =2 .114 , P <0 .05) . The correlation coefficient of LAA maximum ,minimum DLZ and LAA depth between 2D-TEE and RT-3D TEE were 0 .638 ,0 .734 ,0 .647 ( all P =0 .000) ,and 0 .517 ,0 .338 and 0 .591 between RT-3D TEE and LAA CTA ,respectively ( all P <0 .05) ,and 0 .503 ,0 .359 and 0 .610 between 2D-TEE and LAA CTA ,respectively ( all P < 0 .05) . LAA DLZ of LAA angiography ( LAA-A) was ( 22 .6 ± 3 .5 ) mm . Fourty-seven AF patients achieved successful LAA occlusion with Watchman device and the size of closure was ( 27 .5 ± 3 .3) mm . Correlation coefficient between the size of device and LAA maximum DLZ by RT-3D TEE ,2D-TEE and LAA CTA were 0 .693 ,0 .647 ,0 .586 , respectively (all P = 0 .000) . Correlation between LAA-A and the size of device was the closest ( r =0 .914 , P =0 .000) .Conclusions The difference in LAA size measured by TEE and LAA CTA could be significant ,and need to be considered before the LAAC . 2D-TEE , RT-3D TEE and CTA LAA measurements all could play important roles in the selection of proper Watchman device size ,and the maximum DLZ of RT-3D TEE has the closest correlation with Watchman device size .