Comparison analysis of WE and SPAIR technology on breast iShim-IVIM imaging
10.3969/j.issn.1006-5725.2017.17.031
- VernacularTitle:精准频率反转恢复及水激发两种压脂方式在乳腺iShim-IVIM成像中的对比
- Author:
Qiang ZHANG
1
;
Qimin GAO
;
Xuexue ZOU
;
Dejing MA
;
Hu ZHANG
;
Dongjing QIN
Author Information
1. 滨州医学院附属医院放射科
- Keywords:
MRI;
intravoxel incoherent motion;
integrated specific slice dynamic Shim;
breast tumor
- From:
The Journal of Practical Medicine
2017;33(17):2926-2929
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the strengths and weaknesses of SPAIR and WE in iShim-IVIM of breast gland MRI scanning. Methods Twenty patients with breast tumor underwent iShim-IVIM and DCE-MRI scanning by SPAIR and WE technology. SNR ,CNR and ADC of the SPAIR-iShim and WE-iShim imaging were cal-culated and imaging quality were evaluated. Statistical analysis was done afterwards. Results The SNR and CNR of WE technology is higher than those of SPAIR statistically (P<0.05). There is no significant statistical differ-ence in score of the images between SPAIR and WE(P>0.05). The ADC of the center of the tumor has no signifi-cant statistical difference between SPAIR and WE(P>0.05). Conclusion The quality of the images using WE is significantly improved than SPAIR ,but both of them can meet the diagnostic requirements ,and the ADC has no statistical difference as well.