In vivo experiment on treatment of vascular disconnection and defect of extremities with interlocking vascular shunt device
10.7687/j.issn1003-8868.2017.09.020
- VernacularTitle:拉索式血管分流装置治疗四肢血管离断并缺损伤的体内试验
- Author:
Xin WANG
1
;
Li LI
;
Qiang LI
;
Bin XU
;
Kerimulla MAMAT
;
Yun HE
;
Gang XU
;
bo Hai LU
;
Jiang PENG
Author Information
1. 新疆军区总医院全军骨科中心
- Keywords:
interlocking;
vascular shunt;
vascular disconnection;
in vivo experiment;
limb ischemia
- From:
Chinese Medical Equipment Journal
2017;38(9):20-22,33
- CountryChina
- Language:Chinese
-
Abstract:
Objective To execute in vivo experiment on the treatment of vascular disconnection and defect of the extremities with interlocking vascular shunt device.Methods An animal model with vascular disconnection and defect of the extremities was established.The shunt device and silicone tube were used to connect the two ends of the disconnected vessel,and then canalization and thrombosis by the device were compared with them by sutures and silicon tube.Results The shunt device combined with silicon tube gained the same canalization as that by sutures and silicon tube, while had significantly shorter operation time (P<0.05).Conchusion The interlocking shunt device restores efficiently blood supply of vascular disconnection and defect of the extremities,and thus facilitates blood vessel grafting in rear hospitals.