Comparative study on HGB and HCT results by ABL800 blood gas analyzer and BC5800 blood cell analyzer
10.7687/j.issn1003-8868.2017.07.097
- VernacularTitle:ABL800血气分析仪与BC5800血细胞分析仪检测HGB与HCT结果的比较研究
- Author:
zhen Yu XU
1
;
zhong Zhi GAN
Author Information
1. 363000福建漳州,解放军175医院,厦门大学附属东南医院检验科
- Keywords:
analysis of blood gas;
analysis of blood cell;
HGB;
HCT;
blood sample
- From:
Chinese Medical Equipment Journal
2017;38(7):97-100
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the difference between hemoglobin (HGB) and hematocrit (HCT) results by ABL800 blood gas analyzer and BC5800 blood cell analyzer respectively.Methods ABL800 blood gas analyzer and BC5800 blood cell analyzer were used to detect HGB and HCT in 97 samples of heparin-anticoagulated arterial blood and EDTA-3K anticoagulated versus blood from July 17 to 21,2016 as well as 63 samples of heparin-anticoagulated arterial blood from August 4 to 9,2016.Results Almost all the results by ABL800 blood gas analyzer were higher than those by BC5800 blood cell analyzer.Conclusion There're significant differences between ABL800 blood gas analyzer and traditional way when used to detect HGB and HCT.Blood cell analyzer is the preferred choice to detect HGB and HCT,and the results by the blood gas analyzer can be used for references.