Verification and application of standard method for detecting serum fluoride
- VernacularTitle:血清氟测定标准方法的验证及应用
- Author:
Ping, YE
;
Yuan, LIU
;
Jian-chao, BIAN
- Publication Type:Journal Article
- From:Chinese Journal of Endemiology
2001;20(2):140-142
- CountryChina
- Language:Chinese
-
Abstract:
Objective Standard method of testing serum fluoride was tested and verified by certified reference material CRM of free fluo ride component in cattle serum (GBW 09143~09144).The method was suited for fluor ide research to assess its practical value.Methods ISE-standard curve,linear relationship,linear range,sensitive and precision were disired result by CRM.400 serum of fluoride in rats were measured.Results The linear ranges of the method were 0.02~5.0 mg/L,r=0.9999,b=57.8,and test lower limit was 0.013 mg/L.The test value was equal to standard value.Conclusions The standard method is accurate,sensitive,convenient and small sample is used;and it is easy to spread.Thus,it has been of practical value.