Central corneal thickness using EX500 excimer laser workstation
10.3980/j.issn.1672-5123.2014.10.24
- VernacularTitle:EX500准分子激光治疗仪术中测量中央角膜厚度的研究
- Author:
Jiao, CHEN
;
Hua, WANG
;
Shu-Xi, HE
;
Dong-Qiang, LUO
- Publication Type:Journal Article
- Keywords:
central corneal thickness;
EX500;
A -scan;
LASIK
- From:
International Eye Science
2014;(10):1828-1830
- CountryChina
- Language:Chinese
-
Abstract:
AIM: To evaluate the accuracy of central conreal thickness ( CCT ) using EX500 Excimer Laser workstation (EX500) in laser in situ keratomileusis (LASIK) patients.METHODS:The CCT of 120 eyes (63 patients) who had LASIK between January 2013 and June 2013 were measured by A- scan and EX500. Three groups were classified: >550μm, 500 ~550μm, <500μm according the CCT value of A-scan. The CCT were measured again by corneal flap creating by moria SBK microkeratome. The thickness of the corneal bed stroma were measured by A-scan and EX500 after keratomileusis. All outcomes were analyzed with paired t test.
RESULTS: The average preoperative CCT value was 527. 9±34. 3μm measured by A-scan, 528. 5±34. 6μm measured by EX500. There was no significant difference between these two measurements (t=1. 736, P=0. 085). In group which CCT >550μm, the average preoperative CCT value was 571. 4±17. 3μm measured by A-scan, 572.7±15. 7μm measured by EX500. There was no significant difference between these two measurements (t=1. 857, P=0. 072). In group which CCT 500 ~ 550μm, the average preoperative CCT value was 523. 4±13. 1μm measured by A-scan, 524. 2±12. 4μm measured by EX500. There was no significant difference between these two measurements ( t=1. 934, P = 0. 058 ). In group which CCT <500μm, the average preoperative CCT value 484. 5±9.8μm measured by A-scan, 483. 7±8. 9μm measured by EX500. There was no significant difference between these two measurements (t=1. 395, P=0. 174). The average CCT value after corneal flap lifting was 401. 3 ± 34. 2μm measured by A-scan, 393. 4±38. 9μm measured by EX500. There was a significant difference between these two measurements ( t = 6. 669, P = 0. 000 ). The average thickness of the corneal bed stroma value after keratomileusis was 332. 6±38. 3μm measured by A-scan, 307. 3 ± 37. 1μm measured by EX500. There was a significant difference between these two measurements ( t=17. 165, P=0. 000).
CONCLUSION: There is no significant difference between preoperative CCT value measured by A-scan and EX500. After corneal flap lifting and keratomileusis, the CCT value measured by EX500 is smaller than measured by A-scan.