Application of the domestic rebounder tonometer in children
10.3980/j.issn.1672-5123.2014.07.42
- VernacularTitle:国产回弹式眼压计在儿童眼压测量中的应用
- Author:
Xu-Chong, PAN
- Publication Type:Journal Article
- Keywords:
intraocular pressure;
rebound tonometer;
children
- From:
International Eye Science
2014;(7):1322-1324
- CountryChina
- Language:Chinese
-
Abstract:
AlM:To evaluate the clinical application of the domestic rebound tonometer ( RBT ) in measuring intraocular pressure ( lOP) in healthy children.
METHODS:lOP measurement was measured in bilateral eyes of 108 cases with domestic RBT. Then the lOP for the left eyes of the children older than 13 years was obtained with Goldmann applanation tonometry ( GAT) .
RESULTS: Of the 108 children, 100 ( 92. 6%) had lOP measurement both eyes successfully. Of the 100 children, 24 were older than 13 years. The mean lOP values for the left eyes of the 24 children of RBT and GAT were (16. 0± 3. 0) mmHg (1kPa=7. 5mmHg) and (15. 7±2. 8) mmHg, respectively. There was no significant difference between the lOP of the two tonometers (P>0. 05). The lOP of RBT had a good liner relationship with that of GAT ( r=0. 849, P<0. 05). The lOP values were normally distributed for the both eyes. Mean lOP values were (16. 0±2. 8) mmHg and (16. 2 ± 2. 6 ) mmHg for the left eyes and right eyes, respectively. There was no statistically significant difference in mean lOP readings between them, which had a good liner relationship (r=0. 863, P<0. 05).
CONCLUSlON: Domestic RBT is a good instrument in detecting lOP in children and it seems to be very comfortable and easy when performing lOP measurement in children without an anesthetic.