Repeatability and reproducibility of retinal straylight measurement using C-Quant compensation comparison method
10.3760/cma.j.issn.2095-0160.2012.02.017
- VernacularTitle:C-Quant散射光计量仪测量散射光的重复性和再现性研究
- Author:
Jin-hai, HUANG
;
Shi-shi, LIN
;
Qin-mei, WANG
;
Bo, ZHENG
;
Si-si, CHEN
;
Rong-rong, GAO
- Publication Type:Journal Article
- Keywords:
C-Quant;
Straylight;
Myopia;
Laser in situ keratomileusis;
Repeatability;
Reproducibility
- From:
Chinese Journal of Experimental Ophthalmology
2012;30(2):160-163
- CountryChina
- Language:Chinese
-
Abstract:
BackgroundIn an ideal eye there would be no light scattering at all,but the eye media is not optically ideal.Intraocular straylight causes a veil of light and reduction in the contrast of the retinal image and thus decrease the quality of vision.ObjectiveThe present study was to investigate the repeatability and reproducibility of C-Quant straylight meter( Oculus,Germany)in measuring retinal straylight of myopia and post-laser in situ keratomileusis(LASIK) corneas.MethodsThis is a prospective research.The consecutive 35 eyes of 21 myopic patients and 34 eyes of 22 patients who received LASIK were included in this trail.Retinal straylight was measured for 7 times at a period of time and analyzed quantificational to evaluate the repeatability of measurement.Thirty-eight eyes of 19 patients were measured again at 3-7 days for 3 times at a period of time to assess the reproducibility of C-Quant straylight meter.The mean standard difference (SD) and coefficient of variation (CV) were used as the credibility evaluation.This clinical study complied with Helsinki Declaration and the informed consent was obtained prior to the medical procedure.ResultsThe straylight Log(s) of 7 times measurement were all less than 0.95.The mean Log(s) were 0.92±0.12 and 0.93±0.17 respectively in myopia group and post-LASIK group,without statistically significant differences among 7 times measurement( F=0.335,P=0.812;F=1.000,P=0.409).The mean SD for the 7 times measurement was 0.07 Log units.SD and CV increased with the number of measurements.The differences of mean SD and CV between 3 times result and 6 times result were significant different (t =-2.080,P =0.045;t =-2.190,P =0.035 ).No difference was found between different time periods( t =-0.531,P=0.598 ).The difference of the results between two measurements from the same patient was 0.013.ConclusionsC-Quant is a noncontact,noninvade,rapid and convenient method for the measurement of straylight in myopia and post-LASIK eyes due to the high repeatability and reproducibility.