Reproducibility and tolerability of iCare rebound tonometer and its agreement evaluation with Goldmann applanation tonometry
10.3760/cma.j.issn.2095-0160.2011.01.019
- VernacularTitle:iCare回弹式眼压计测量的重复性及与Goldmann压平式眼压计测量值间的一致性研究
- Author:
Shun-hua, ZHANG
;
Xiao-li, LIU
;
Jia-liang, ZHA
- Publication Type:Journal Article
- Keywords:
Intraocular pressure;
ICare rebound tonometer;
Reproducibility;
Correlation;
Goldmann applanation tonometer
- From:
Chinese Journal of Experimental Ophthalmology
2011;29(1):73-77
- CountryChina
- Language:Chinese
-
Abstract:
Background Goldmann applanation tonometry (GAT) is a gold standard of intraocular pressure measurement.But its utilization iS limited because of its complexity and high requirement of cooperation.iCare rebound tonometer (iCare RBT) is a new type of applanation tonometry,and its accuracy and safety in clinical application need to be evaluated.Objective Present study was to investigate the reproducibility and tolerability of iCare RBT and its measurement agreement with GAT over a wide intraocular pressure (IOP) range. Methods The IOP were measured in bilateral eyes of 36 glaucoma and glaucoma suspect patients by 2 examinees with iCare RBT at the 1-minute interval to assess the interobserver reproducibility.Then the IOP of both eyes from 56 Subjeets and other 52 subjects were separately measured twice for each by two operators with iCare RBT for the evaluation of intraobserver reproducibility.Finally.IOP of 182 eyes of 92 glaucoma and glaucoma suspect patients was obtained by examiner 2 with RBT first and examiner 1 with GAT subsequently at a 2.minute interval in a masked fashion to perform an agreement evaluation of two readings by using Bland-Ahman method.The tolerance of subjects to iCare RBT measurement were surveyed.Oral informed consent was obtained prior to the IOP measurement. Results Concerning the iCare RBT readings.interobserver correlation coefficients were 0.937 in the right eye and 0.887 in the left eye.Intraobserver correlation coefficients of examiner 1 were 0.986 in the left eyes and 0.969 in the fight eyes.And those of examiner 2 were 0.990 and 0.979.Mcan values of iCare RBT readings and GAT were(18.74±8.36)mmHg and(19.33±8.20)mmHg and the mean difference values(iCare-GAT)was(-0.59 4±2.60)mmHg with the 95%confidence interval of -5.80-4.60 mmHg.The correlation coefficient between two modalities of IOP measurement WaS 0.95 1.No severe pain and discomfort were complained in all the subjects during or after measurement of iCare RBT. Conclusion iCare RBT has good interobserver and intraobserver reproducibility and good tolerance.It was proved that this is a good correlation between iCare RBT readings and GAT readings.