Oto-drill (A prototype surgical drill unit)
- Author:
Paul Jansen T Alcaraz
;
Francis V Roasa
- Publication Type:Journal Article
- Keywords:
OTO-DRILL SURGICAL DRILL
- MeSH:
SURGERY OTOLARYNGOLOGY
- From:
Philippine Journal of Otolaryngology Head and Neck Surgery
2005;20(1-2):55-58
- CountryPhilippines
- Language:Filipino
-
Abstract:
OBJECTIVE: To design and produce a high quality, inexpensive surgical drill using locally available electronics. DESIGN: Descriptive (Instrument Design) SETTING: University based tertiary hospital MATERIALS AND METHODS: The surgical drill apparatus will use high quality electronics readily available locally which will make it cheaper than the other available units. This prototype unit was designed and tested by the author in collaboration with the Santo Tomas University Hospital Biomedical Engineers. The drill units schematics and specifications where patterned from the RAM Microtorque II drill unit which is produced in the U.S.A. RESULTS: The prototype is a cost effective alternative to the expensive drill units available in the market. It was given a 24 hour durability test which it passed with no loss of power or drilling efficiency proving that it can be used for major surgeries especially in the field of Otolaryngology and Neurosurgery. CONCLUSION: The OTODRILL is a cheap alternative and of comparable quality to expensive surgical drills available in the market which can be used in major surgeries in the field of Otolaryngology and other surgical fields. (Author)