Sterile repositionable hemostasis clipping device in combination with high frequency electric snare in resection of colorectal polyps with wide and long peduncle
10.3969/j.issn.1007-1989.2016.08.023
- VernacularTitle:可旋转重复开闭软组织钛夹联合高频电圈套电凝切除治疗大肠粗长蒂息肉的临床安全性观察
- Author:
Xiaorui WANG
;
Xiaorong WANG
;
Yan ZHANG
;
Liqing YANG
- Keywords:
sterile repositionable hemostasis clipping device;
colorectal polyps;
electric resection of high frequency electric snare and coagulation
- From:
China Journal of Endoscopy
2016;22(8):99-101
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss the efficacy and safety of sterile repositionable hemostasis clipping device in combination with high frequency electric snare in polypectomy of colorectal polyps with wide and long peduncle. Methods From January 2014 to December 2015, 21 cases of colorectal polyps with wide and long peduncle (diameter greater than 2 cm) in endoscopic treatment under electronic colonoscopy. We used 1 ~ 2 hemostatic clips to clip colorectal polyps roots, then used electric resection with high frequency electric snare electric coagulation. Postoperative bleeding, perforation were observed follow-up. Results 24 polyps in 21 cases were removed one-time successfully. Stump errhysis in 1 case, hot biopsy forceps is given to deal with local wound followed by Olympus Clip HX-610-135L EZ titanium clip. There was no complication such as bleeding and perforation in 3 to 6 months after the operation. In the colonoscopy examination, recurrence of polyps were not found in the original polyp resection site. Conclusion Sterile repositionable hemostasis clipping device in combination with high frequency electric snare in polypectomy of colorectal polyps with wide and long peduncle is safe and effective, without bleeding or perforation.