Decentration and tilt of IOL after capsule tension ring implantation observed by ultrasonic biomicroscope
10.13389/j.cnki.rao.2017.0142
- VernacularTitle:超声生物显微镜观察囊袋张力环应用后人工晶状体的偏心量和倾斜度
- Author:
Yutong HAN
;
Binchuan CHEN
;
Guangju ZHU
;
Yifan XU
;
Fang LEI
- Keywords:
cataract;
high myopia;
ultrasonic biomicroscope;
intraocular lens;
capsular tension ring
- From:
Recent Advances in Ophthalmology
2017;37(6):562-565
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss the effects of capsular tension ring after phacoemulsification combined with IOL implantation on tilt and decentration of IOL in high myopia patients with cataract by ultrasonic biomicroscope (UBM).Methods A total of 36 cases (40 eyes) with high myopia and cataract underwent phacoemulsification combined with IOL implantation were chosen.The average axial length was 26.88 mm.The patients were divided into implant group (19 eyes,the capsular tension ring was implanted) and control group (21 eyes,the routine surgery was performed).The patients were examined by conventional slit lamp,and the best corrected visual acuity (BCVA)was measured at pre-operation and postoperative 6 months.Tilt and decentration were measured horizontally and vertically,and total tilt and decentration were calculated by geometry method.Results The postoperative BCVA in the two groups were all better than the pre-operation,there was no statistical difference in the preoperative and postoperative BCVA between two groups (all P > 0.05).The horizontal,vertical and total decentration the implant group were (0.15 ± 0.07) mm,(0.30 ± 0.40) mm,(0.11 ±0.02)mm,respectively,which in the control group were (0.26 ± 0.19)mm,(0.32 ±0.60) mm,(0.24 ± 0.97) mm,respectively.The horizontal,vertical and total tilt in the implant group were 0.02° ±0.11°,0.70° ±0.25°,0.21° ±0.74°,respectively,which in the control group were 0.11 ° ± 0.31 °,1.09° ± 0.20°,1.24° ± 0.97°,respectively.There were statistical differences in the horizontal,total tilt and decentration between two groups (all P < O.05),but no statistical difference in the vertical tilt and decentration (P > 0.05).Conclusion The capsular tension ring can stable the IOL position after surgery in high myopia and cataract patients.