Comparison study of readout-segmented EPI and single-shot EPI in DWI of orbit in healthy adults
10.13929/j.1003-3289.201609023
- VernacularTitle:RS-EPI与SS-EPI序列在眼眶弥散加权成像中的对照研究
- Author:
Jun LIU
;
Xiaoquan XU
;
Feiyun WU
- Keywords:
Orbit;
Diffusion magnetic resonance imaging;
Echo planar imaging
- From:
Chinese Journal of Medical Imaging Technology
2017;33(6):872-875
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the feasibility of the applicatoin of readout-segmented echo-planar imaging (RS-EPI) at 3.0T MR of the orbit DWI and compared with single-shot echo-planar imaging (SS-EPI).Methods Forty-two volunteers underwent both standard SS-EPI and RS-EPI DWI of the orbit at a 3.0T MR unit.Two sets of DWI images were independently qualitatively scored in the number of distinguishable normal structures,fat suppression,ghosting artifact and overall image quality.SNR of the vitreous body,geometric distortion ratio (GDR) in both anterior-posterior (AP) and right-left (RL) direction,and ADC value of the vitreous body and pons were also quantitatively calculated and compared between two sets of DWI images.The statistical analysis was performed.Results For qualitative assessment,the RS-EPI was superior to SS-EPI on the number of distinguishable normal structures (P=0.009 0),ghosting artifact (P<0.000 1),and overall image quality (P<0.000 1),while no significant difference was found on fat suppression (P=0.753 9).For quantitative assessment,RS-EPI had significantly lower GDR than the SS-EPI in both AP and RL direction (P=0.001 4,0.001 7).The SNR in RS-EPI was significantly lower than that of SS-EPI (P=0.004 0).Meanwhile,there was no significant difference of ADC on vitreous body and pons between RS-EPI and SS-EPI (P=0.143 8,0.126 2).Conclusion The RS-EPI can provide better image quality than SS-EPI protocol in orbital DWI at 3.0T MR imaging.