Abnormal results analysis and quality control of biological monitoring by pressure steam sterilizer
10.7687/j.issn1003-8868.2017.05.110
- VernacularTitle:压力蒸汽灭菌器生物监测结果异常分析
- Author:
Mingxiu LIU
;
Ling WANG
;
Hongling FAN
- Keywords:
pressure steam sterilizer;
biological monitoring;
bioindicator;
sterilization effect;
quality control
- From:
Chinese Medical Equipment Journal
2017;38(5):110-112
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the causes for the biological monitoring abnormal results of the pressure steam sterilizer to eliminate non-standard operation and hidden risks during biological monitoring.Methods Totally 34 secondary or above medical institutions in Chongqing were selected randomly,and questionnaire survey was carried out over biological monitoring abnormal results and main influencing factors of the pressure steam sterilizer.Results Of the 126 abnormal results,there were 63 positive results (accounting for 50%) and the main influencing factors included expiration,aged seal ring,low-efficiency vacuum pump,nonstandard loading,bioindicator reproduction before sterilization,false positive result,polluted moist incubator and etc;There were 35 cases of results related to fast reader (accounting for 27.78%) and the main factors consisted of failure of reading light,improper input of bioindicator,dusts in the reading hole and etc;There were 18 cases of bioindicator discoloration before culture (accounting for 14.28%) and the main factor was reproduction occurred before the gemma was inactivated;There were l0 cases of nondiscoloring in case of positive control (accounting for 7.94%) and the main factors included non-crushed indicating mediostratum cultured in the incubator,insufficient shaking after crushing mediostratum and etc.Conclusion The main factors for abnormal biological monitoring of the sterilizer involve in equipment,personnel and material,and the quality control has to be executed from the three aspects.