Image Quality Assessment of Brain CT with Different Scanning Modes
10.3969/j.issn.1005-5185.2017.06.005
- VernacularTitle:头颅CT不同扫描方式的图像质量分析
- Author:
Xiaoyi LIU
;
Weiwei QI
;
Zhuo LIU
;
Lei CHEN
;
Nan HONG
- Keywords:
Tomography,X-ray computed;
Tomography,spiral computed;
Brain;
Radiation dosage;
Quality control
- From:
Chinese Journal of Medical Imaging
2017;25(6):418-421
- CountryChina
- Language:Chinese
-
Abstract:
Purpose To analyze the image quality of brain CT with 256-slice wide detector axial scanning mode,routine axial scanning mode and spiral scanning mode,and to provide a more effective brain CT examination method for patients.Materials and Methods The prospective study was conducted on 90 patients accepting routine brain CT examination,and they were randomly divided into three groups.CT examination with 160 mm axial scanning mode,40 mm axial scanning mode and 40 mm spiral scanning mode were respectively conducted using GE Revolution CT.The scanning condition was adjusted to remain constant radiation dose,and then the image quality was analyzed.CT attenuation of gray and white matter,contrast-to-noise ratio (CNR) of white-gray matter and image noise of the three scanning modes were compared.Subjective scoring on image quality of the three scanning modes was also performed.Results On body lateral cerebral ventricle level,there were no significant difference in CT attenuation of gray and white matter and CNR (P>0.05).On centrum semiovale level,the CT attenuation of gray matter [(31.71 ± 1.82) HU],white matter [(22.97± 1.50) HU] and CNR 2.05±0.42 of 160 mm axial scanning mode was significantly different from the other two scanning modes (F=26.74,47.16 and 3.85,P<0.05).On centmm semiovale level,image noise of 160 mm axial scanning mode was lower than the other two kinds of scanning methods (F=6.31,P<0.05),in the basal ganglia and posterior fossa there were no statistically significant differences in the image noise between the three scanning modes (P>0.05).The subjective score of the three scanning modes all met the diagnostic requirements,and there was no significant difference (P>0.05).The effective dose and scanning time of 160 mm axial scanning mode was lower than those of the other two scanning modes,and the X-ray utilization was higher.Conclusion 160 mm wide detector axial scanning mode is more suitable for brain CT scan,and it can be used as the preferred scanning mode in the emergency and among non-cooperative patients.