Practice and discussion of the quality control for injection pump
10.3969/J.ISSN.1672-8270.2017.08.008
- VernacularTitle:注射泵质量控制实践及探讨
- Author:
Yufeng CHANG
;
Hongli PEI
;
Guangquan LIU
- Keywords:
Injection pump;
Quality control;
Flow test;
Blocking pressure
- From:
China Medical Equipment
2017;14(8):29-31
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To discuss the quality control method of injection pump and analyze the factors that affect the quality control of injection pump.Methods: Flow error and blocking alarm error were tested by the INFUTEST 2000E dual-channel infusion pump analyzer according to the national calibration specifications.Results: The syringe specifications, injection speed and injection solvent could influence the test for accuracy of injection pump. The lower injection speed set, the bigger flow test error was, while the smaller the blocking alarm test error was. On the other hand, the higher the injection speed set, the smaller the flow test error was and the bigger the blocking alarm test error was. Besides, the higher qualified rate of flow test was, the lower the qualified rate of blocking alarm test was.Conclusion: The quality control of injection pump should abide by the principle of controllable risk, and should not pursuit too much accuracy.