A novel double-ring aortic valve stent: its development and experimental study
10.3969/j.issn.1008-794X.2017.04.014
- VernacularTitle:新型双环状主动脉瓣支架研制及实验研究
- Author:
Xiang CHEN
;
Tong KAN
;
Guojun CHU
;
Ben ZHANG
;
Feiyu WANG
;
Jian ZHANG
;
Yongwen QIN
;
Zhijun ZHU
;
Danning WU
- Keywords:
transapical route;
transcatheter aortic valve implantation;
aortic valve insufficiency
- From:
Journal of Interventional Radiology
2017;26(4):344-349
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss the feasibility and effectiveness of transcatheter implantation of double-ring aortic valve stent through puncturing the tip of the heart under thoracotomy.Methods A novel double-ring aortic valve stent was independently designed by the authors.Three healthy goats were selected for this study.A small incision on the left anterolateral thoracic wall was made to expose the cardiac apex,than the puncturing of the left ventricular apex was performed to establish the working pathway.Guided by fluoroscopy,along a hard guide wire a double-ring aortic valve stent was inserted through a 22-French sheath to the site above the aortic valve.By utilizing the opened outer ring of the stent,the double-ring aortic valve stent was accurately placed at the bottom of the aortic valve sinus.Then,the balloon was inflated and the stent was released to substitute the original aortic valve of the experimental goat.The experiment results were evaluated immediately after the procedure.Results Transcatheter aortic valve implantation (TAVI) was successfully accomplished in all the three experimental goats.DSA was performed immediately after the procedure and anatomy evaluation indicated that the position of the implanted artificial aortic valve was satisfactory,which could replace the work of original valve.Conclusion It is technically feasible and clinically effective to use this novel double-ring aortic valve stent to perform TAVI through transapical route by puncturing the left ventricular apex.