The performance evaluation on 3 hemoglobin analyzers
10.13303/j.cjbt.issn.1004-549x.2017.07.071
- VernacularTitle:三种血红蛋白分析仪的性能评价
- Author:
Bei XU
;
Jianhao YANG
;
Xiaoyan ZHOU
;
Yingjie QIU
- Keywords:
performance evaluation;
hemoglobin;
precision;
linearity
- From:
Chinese Journal of Blood Transfusion
2017;30(7):846-848
- CountryChina
- Language:Chinese
-
Abstract:
Objective To carry out the performance evaluation on 3 hemoglobin analyzers-URIT-12,HemoCue Hb301 and CompoLab TS.Methods Verification of inter-run precision,intra-day precision,and correlation of the 3 analyzers using samples with different concentrations.Results The lower values of inter-run precision of URIT-12,HemoCue Hb301 and CompoLab TSoft were 1.46%,1.45% and 1.32%,respectively,and the median values were 1.43%,0.56% and 1.49%,respectively.The lower values of intra-day precision were 1.86%,1.62% and 1.32% respectively for URIT-12,HemoCue Hb301 and CompoLab TSoft,with the median values of 1.45%,1.95% and 1.78% and the higher values of 1.86%,1.74% and 1.81%.Based on the results of Beckman-Coulter AcT5diff automatic blood cell analyzer,linear regression equation and correlation coefficient was calculated and expressed as y=0.960 2x+3.201 3,R2=0.973 4;y=1.112 1x-6.767 9,R2 =0.990 7;y=1.031 5x+0.634 7,R2=0.980 3,respectively.Conclusion The 3 hemoglobin analyzers have better inter-run precision and intra-day precision as well as good correlation with automatic blood analyzer,and thus can be applied in physical examinations before blood donation.