Determination of Total Chloride in Oral Rehydration Salt Powder (Ⅲ) by Ion Chromatography
10.3969/j.issn.1008-049X.2017.09.049
- VernacularTitle:离子色谱法测定口服补液盐散(Ⅲ)中总氯的含量
- Author:
Jianyong HU
;
Tingshun LUO
;
Lihua ZHANG
- Keywords:
Ion chromatography;
Oral rehydration salt powder (Ⅲ);
Total chloride
- From:
China Pharmacist
2017;20(9):1689-1691
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To establish a method for the determination of total chloride in oral rehydration salts powder (Ⅲ) by ion chromatography (IC). Methods:The analysis was performed on an IonPac AS 14A (250 mm × 4 mm) column with an IonPac AG 14A (50 mm × 4 mm) guard column and 9mmol·L-1 Na2 CO3 solution as the mobile phase. The flow rate was 1. 0 ml·min-1 with isocratic elution and the sample size was 25 μl. The suppressor was AERS 500(4mm) with 45mA suppression current, and an ECD was used for the detection. The quantity of chloride ion ( Cl-1 ) was calculated by the peak area with an external standard method. Re-sults:The linear range of Cl-1 was 0. 05-20. 00 mg·L-1(r=0. 9999);the recovery was 98. 89%(RSD=0. 56%, n=6). Conclu-sion:The results demonstrated that the method has the advantages of high sensitivity, simple sample pretreatment, promising accuracy and good reliability, which is suitable for the determination of total chloride in oral rehydration salt powder (Ⅲ) .