Simultaneous Determination of 6 Residual Organic Solvents in Aprepitant Raw Material by Headspace Capillary GC
10.6039/j.issn.1001-0408.2017.24.33
- VernacularTitle:顶空毛细管气相色谱法同时测定阿瑞匹坦原料药中6种有机溶剂的残留量
- Author:
Ruifeng ZHENG
;
Chen YANG
;
Fengying REN
;
Hongqian JIA
;
Lin RAN
;
Qihui QIN
;
Xiaojun GOU
;
Ju FENG
- Keywords:
Headspace capillary gas chromatography;
Aprepitant;
Organic solvent;
Residual
- From:
China Pharmacy
2017;28(24):3426-3429
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To establish a method for simultaneous determination of 6 residual organic solvents in aprepitant raw material as methanol,ethanol,acetone,isopropyl alcohol,methyl tert-butyl ether and tetrahydrofuran.METHODS:Headspace capillary gas chromatography was adopted.The determination was performed on DB-624 capillary column using temperature programming.The temperature of injector port was 180 ℃,and flame ionization detector was used with temperature of 260 ℃.Nitrogen was used as carrier gas with flow rate 3.0 mL/min.The spilt ratio was 5 ∶ 1,and head-space injection volume was 1.0 mL.The head-space equilibrium temperature was set at 80 ℃,and equilibrium time was 40 min.RESULTS:The linear ranges of methanol,ethanol,acetone,isopropyl alcohol,methyl tert-butyl ether,tetrahydrofuran were 6.052-605.232 μ g/mL (r=0.999 9),9.987-998.718 μg/mL(r=0.999 9),9.998-999.768 μg/mL(r=0.999 8),9.986-998.634 μg/mL(r=0.999 9),9.991-999.090 μg/mL (r=0.999 7),1.461-146.133 μg/mL(r=0.999 5),respectively.The limits of quantitation were 1.782 1,2.079 0,0.749 8,1.777 8,0.223 1,0.607 0 μg/mL;the limits of detection were 0.594 0,0.693 0,0.249 9,0.592 6,0.074 4,0.202 3 μg/mL,respectively.RSD of precision test was lower than 2.0%.Only acetone and isopropyl alcohol were detected in stability test and reproducibility tests,RSD<2.0%.Their recoveries were 99.34-100.75% (RSD=0.52%,n=9),98.20%-100.24% (RSD=0.69%,n=9),98.07%-100.07% (RSD=0.84%,n=9),99.86%-101.32% (RSD=0.58%,n=9),97.87%-104.02% (RSD=2.13%,n=9),98.26 %-100.58 % (RSD =0.75 %,n =9),respectively.CONCLUSIONS:The established method is simple,accurate and reproducible,and can be used for simultaneous determination of 6 residual organic solvents in aprepitant raw material.