Study on endotoxin quantitative detector
10.3969/j.issn.1671-8348.2017.11.021
- VernacularTitle:内毒素定量检测仪的研究
- Author:
Tianyu XIANG
;
Huilai WANG
- Keywords:
endotoxins;
piezoelectric quartz crystal;
sensor;
limulus reagent;
eigenvalue
- From:
Chongqing Medicine
2017;46(11):1506-1507,1510
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop a quantitative,highly sensitive,low detection limit and fast endotoxin(ET) biosensing detection system by utilizing the piezoelectric crystal's damping effect(non-mass effect) in the liquid phase environment.Methods The sensor probe of better performance (eliminating the quality load effects and improving the liquid solid coupling effect) was obtained by using the smoothing processing and hydrophilic processing of a quartz crystal sensor probe surface,thus the stability and repeatability of measurement were increased and the low detection limit of measurement was decreased.The whole measurement system was made into a miniature and automated endotoxin detection system by using CPLD device and LabView software to realize the automatic collection,and analysis and results display of measurement data.Results The good linear relationship existed between different concentrations of endotoxin and the stable value of frequency shift.Conclusion By the established data of frequency shift,the endotoxin concentration can be obtained accurately and conveniently.