According to the platelet of Chinese people biological reference interval change the microscope review rules
10.3969/j.issn.1673-4130.2016.09.014
- VernacularTitle:根据中国人群血小板生物参考区间修改及验证显微镜复检规则
- Author:
Feng WANG
;
Zhen WANG
;
Yanjing WANG
;
Fenghua ZHANG
- Publication Type:Journal Article
- Keywords:
hematology analyzer;
platelet aggregation;
pseudothrombocytopenia
- From:
International Journal of Laboratory Medicine
2016;37(9):1190-1191,1194
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss change the platelet counts by microscope rule by the platelet of Chinese people biological refer‐ence range in this laboratory blood analysis requirement .Methods Retrospectively analyzed according to the platelet of Chinese people biological reference range change the lab Sysmex XE‐5000 automatic blood cell analyzer platelet review rules .Change the platelet counts by microscope rules ,the differences between the microscopy rate and the microscopic examination of the platelet ag‐gregation rate .Results Before change the platelet counts by microscope rules the microscopy rate was 1 .94% .After change the platelet counts by microscope rules the microscopy rate was 3 .99% ,after change the platelet counts by microscope rules pushing blood piece rate increased significantly ,the difference was statistically significant(P<0 .05) .Conclusion According to the platelet of Chinese people biological reference range change the lab platelet review rules .We can increase the rate of microscopy ,and effec‐tively avoid the pseudothrombocytopenia and accurately guide clinical treatment .