Effect of Replacement of New Reverse Osmosis Film of Water Purification System to Clinical Biochemistry
10.3969/j.issn.1671-7414.2014.06.040
- VernacularTitle:实验室纯水系统更换反渗透膜对生化检验的影响
- Author:
Jihao MO
;
Xianru LIU
;
Shaoxia LI
- Publication Type:Journal Article
- Keywords:
triglyceride;
purified water;
reverse osmosis film;
clinical biochemistry
- From:
Journal of Modern Laboratory Medicine
2014;(6):124-126,130
- CountryChina
- Language:Chinese
-
Abstract:
Objective To test if the quality of produced water could come up to the standards of clinical biochemistry,and e-valuate the effect of the produced water after replacement of a new reverse osmosis film of the laboratory water purification system.Methods The ion concentration of the produced water was tested by ohmmeter in Beckman Coulter supporting TianChuang water purification system TCHS-RO/100A and HM COM-100 conductivity meter.The biochemical index ALT,AST,TBIL,ALP,GGT,TP,ALB,BUN,CR,UA,GLU,TC,TG,HDL,LDL,CK,LDH,P,Ca of mixed serum were tested 20 times by AU680 under the condition of normal purified water(control group),and the same operation were repeat-ed after changing to newly produced water(experimental group)to evaluate the effect of newly produced water to biochemi-cal analysis.The triglyceride (TG)reagent blank was tested 3times/d for 5 days to evaluate the glycerin level in the pro-duced water.Results The conductivity of purified water was reduced to (0.1±0.0)μS/cm 30 min after starting up by HM COM-100,and the electric resistance of purified water remained stable at 18.25MΩcm 50min after starting up.During the bi-ochemical analysis by AU680,the value of TG in control group=1.04 mmol/L,in experimental group=21.39 mmol/L,the differences showed statistical significance (P<0.01),but the bias of the others were limited in 5%,CV<1/4CLIA’88,and differences showed no statistical significance (P>0.05).The reagent blank was stable at A660nm=0.015 6±0.000 4 in the fifth day.Conclusion After produced approximately 2 800 L purified water in the system,glycerine could be cleaned up in produced water.It is needed to continuously test the quality of produced water to evaluate the effect of which to clinical bio-chemistry after replacement of a new reverse osmosis film.