Preliminary study of radiation dosage in the lower limb factures surgery with single source dual energy CT
10.3760/cma.j.issn.1008-6706.2015.04.008
- VernacularTitle:单源能谱CT在下肢骨折术后辐射剂量的初步研究
- Author:
Sensen YAN
;
Yang LI
;
Min LUO
;
Xiuya CEN
- Publication Type:Journal Article
- Keywords:
Fractures,Bone;
Tomography,X-Ray Computed;
Radiation Dosage
- From:
Chinese Journal of Primary Medicine and Pharmacy
2015;22(4):500-503,641
- CountryChina
- Language:Chinese
-
Abstract:
Objective To assess the clinical application value of single-source dual energy CT(SSDE),by measuring the multi-slice CT(MSCT) and single-source dual energy CT(SSDE) lower limb fractures image noise and contrast noise ratio(CNR),comparing the difference in the radiation dosage.Methods 39 cases of lower extremity fractures by clinical diagnosis were included in the study.Preoperation or postoperation the patients underwent MSCT scans,postoperation or postoperative follow-up SSDE CT scans were conducted,two sets of images were achieved.We measured and compared conventional CT group with 130 key energy spectrum group (prior studies showed the elimination of metal artifacts,images,subjective scoring the highest photon energy level) no significant artifacts cancellous bone area,compared the image background noise and contrast noise ratio (CNR),and radiation dosage,using paired sample t-test.Results Two sets of data for the same patient before and after surgery or after review of images,which body mass index (BMI) was negligible.The CTDIvol value of the conventional MSCT was (7.94 ± 1.69) mGy,the dual-energy CT was (7.04 ± 0.97) mGy,with significant difference (t =2.71,P < 0.05) ; The image noise of the conventional MSCT was (44.36 ± 15.66) Hu,the dual-energy CT was (15.04 ± 8.23) Hu,with significant difference (t =15.26,P < 0.05) ; The CNR of the conventional MSCT was (3.96 ± 1.78),the dual-energy CT was (4.62 ± 2.80),with significant difference (t =-2.14,P < 0.05).Conclusion Compared with conventional MSCT,single source dual-energy CT imaging provides improved image CNR,obviously eliminate metal artifacts,to better show the complex structure for lower limb fracture surgery,and reduction in radiation dosage.