Compensation research on energy response for the semiconductor personal dosimeter
10.3969/J.ISSN.1672-8270.2015.03.002
- VernacularTitle:半导体型个人剂量计能量补偿的研究
- Author:
Xiangming OU
;
Shian ZHAO
;
Yanqiu DING
- Publication Type:Journal Article
- Keywords:
Semiconductor;
Personal dosimeter;
Pb/Sn additional filters;
Energy compensation
- From:
China Medical Equipment
2015;(3):4-6,7
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To complete experiments of the energy response compensation for three semiconductor personal dosimeters in order to meet the requirements of IEC and National Standard.Methods: In Beijing SSDL, the semiconductor detectors were covered by the different thicknesses of Pb and Sn additional filters with several diameter holes to compensate the dosimeters’ energy response.Results: At the 50 keV-Cs-137 energy range, the measuring error of dosimeters is met within 30% of energy response.Conclusion:By the compensation of combined Pb and Sn filters with a hole, the semiconductor personal dosimeters can be used for radiation monitoring.