Analysis on hospital medical equipment electrical safety testing
10.3969/J.ISSN.1672-8270.2015.07.009
- VernacularTitle:医院医疗设备电气安全检测分析
- Author:
Huawei ZHANG
;
Hang QIN
;
Hongbing JIANG
- Publication Type:Journal Article
- Keywords:
Medical equipment;
Electrical safety;
Grounding resistance;
Earth resistance;
Leakage current
- From:
China Medical Equipment
2015;(7):25-26,27
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To study the electrical safety of multi-parameter monitor and defibrillation monitor, and guarantee the safe and effective operation by through the medical equipment electrical safety performance in our hospital.Methods: To test 101 different models of multi-parameter monitor and defibrillation monitor electrical safety.Results: The one-time pass rate of 96%, and the two-time rate was 98%, also one multi-parameter monitor and defibrillation monitor hadn’t meet electrical safety performance after repairing.Conclusion: The equipments that don’t meet electrical safety performance prohibit using for clinical purposes; To ensure the safety of medical equipment and life safety of medical workers and patients by through developing the testing of the medical equipment electrical safety performance in our hospital.