Performance verification of portable Celercare M1 analyzer in CRP determination
10.3969/j.issn.1673-4130.2015.01.016
- VernacularTitle:便携式 Celercare M1分析仪 CRP 检测的性能验证
- Author:
Yonghui DAI
;
Xingguo CHEN
;
Zhanhui WANG
- Publication Type:Journal Article
- Keywords:
C-reactive protein;
performance evaluation;
Celercare M1 analyzer
- From:
International Journal of Laboratory Medicine
2015;(1):39-41,44
- CountryChina
- Language:Chinese
-
Abstract:
Objective To verify the analytical performance of the portable Celercare M1 analyzer in the C-reactive protein(CRP) quantitative determination.Methods According to the standard program of the American Clinical and Laboratory Standards Insti-tute (CLSI),The portable Celercare M1 analyzer was performed the verification on the CRP detection in the aspects of the linearity range,precision,accuracy,interference test and method comparison.Results The CRP detection showed good linearity in the range of 0.5-200 mg/L and the regression equation was Y =0.979X +0.456.The within-run coefficients of variation(CV)of samples with low and high CRP levels were 5.9% and 2.0%,respectively.The between-day CV of samples with low and high CRP levels were 6.3% and 2.2%,respectively.The accuracy of the assay was admirable,and bias of CRP results at low and high levels were 2.3% and 0.7%,respectively.No significance interference was observed when serum bilirubin ≤340 μmol/L,triglyceride ≤10 mmol/L and hemoglobin ≤8 g/L in this method.The regression equation of Celercare M1 analyzer(Y )and the IMMAGE 800 (X) for detecting CRP was Y =0.944X +0.206,R 2 =0.996(P <0.05).The relative bias of CRP results in three medical decision levels (3,10,100 mg/L)was 1 .0%,4.0% and 5.0%,repectively,which was lower than 1/2 of total error allowance.Conclusion The linearity,precision,accuracy and interference of the Celercare M1 analyzer in CRP determination are admirable and can be accepted in clinical practice.