Study of optimization of FLAIR pulse sequence for image quality
10.3969/j.issn.1002-1671.2014.11.035
- VernacularTitle:FLAIR 序列参数优化对图像质量的研究
- Author:
Wei SUN
;
Xiuzhong YAO
;
Hao FENG
;
Caizhong CHEN
;
Mengsu ZENG
- Publication Type:Journal Article
- Keywords:
magnetic resonance imaging;
fluid-attenuated inversion recovery(FLAIR);
brain
- From:
Journal of Practical Radiology
2014;(11):1906-1909,1954
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the effect on image quality and scanning time by changing parameters and using different scan-ning technique for the clinical application of axial FLAIR pulse sequence.Methods Seven different sets of parameters of T2-FLAIR Pulse Sequences were performed on thirty patients at a 3.0 T MR scanner.Group A was scanned with a routine sequence for refer-ence.Group B,reducing the matrix to 128.Group C,increasing the turbo factor parameter to 24.Group D,setting the contatena-tion to 1.Group E setting the Accel.factor PE to 4.Group F,using a BLADE-FLAIR sequence.Group G,using a HASTE-FLAIR sequence.The SNRs of all images were statistically analyzed by ANOVA among 7 sequences.The image quality of all images was statistically analyzed by Kruskal-Wallis test among 7 sequences.Results The SNR of seven groups of images were 54.69 ±8.12, 1 57.57±25.61,44.56±7.37,38.48 ± 9.32,44.1 9 ±8.79,45.60 ± 6.92,105.20 ± 18.25,respectively(F = 266.368,P =0.000). The scores of image quality were 3.87±0.09,2.50±0.12,3.72±0.1 6,3.10±0.10,1.88±0.10,3.92±0.07,2.08±0.10,re-spectively and the differences among groups were statistically significant(χ2 =1 93.361,P =0.000).Conclusion Suitable turto factor and BLADE technique can be the best option for FLAIR sequence in the head.HASTE-FLAIR sequence can act as the secondary op-tion for the uncooperative patients.