The establishment of x-ray environmental standard calibration quality
10.3969/J.ISSN.1672-8270.2014.01.003
- VernacularTitle:X射线环境水平标准装置的研究与建立
- Author:
Xiangming OU
;
Shian ZHAO
;
Yanqiu DING
- Publication Type:Journal Article
- Keywords:
X-ray;
Environmental level;
Radiation qualities;
Standard calibration field;
Half value layer;
Effective energy
- From:
China Medical Equipment
2014;(1):8-10,11
- CountryChina
- Language:Chinese
-
Abstract:
Objective:According to the requirement of ISO and National Standard, new X-ray environmental standard for the calibration of environmental dosimeter was established in Beijing SSDL. Methods:Half value layer and air kerma dose-rate was determined in the X-ray environmental qualities by a standard dosimeter. Results: The technical requirement of ISO and National Standard are met within 2% for the X-ray environmental standard qualities in Beijing SSDL. Conclusion:The X-ray standard radiation field can be used for the calibration of environmental dosimeters and experiment of research.