Comparison of Performances of Three Kinds of Direct Current Power Source in the Application of Electrical Impedance Tomography(EIT)
- VernacularTitle:3种直流电源在电阻抗成像应用中的性能比较研究
- Author:
Jieshi MA
;
Xuetao SHI
;
Canhua XU
;
Meng DAI
;
Xiuzhen DONG
- Publication Type:Journal Article
- Keywords:
direct current power supply;
spectrum;
SNR;
EIT
- From:
Chinese Medical Equipment Journal
2003;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To provide the hardware system of EIT with an excellent direct current(DC) power source. Methods The spectrums of the outputs of DC power source and the driving source were analyzed by means of Spectrum Analyzer- E4408B when EIT system was respectively powered by battery, linear power source and switching power source. Based on the numbers and power levels of the spurs of the three power sources and the numbers and power levels of the interferences within the driving source, the influence of the three power sources on the error of measuring was estimated. Meanwhile the EIT system did data acquisition, the overall signal-to-noise-ratio(SNR) of the three power sources at the 18 frequencies were calculated and the influence of the three power sources on the error of measuring was judged. Results With respect to the numbers and power levels of the interferences within the driving source when EIT was powered respectively by the three power sources, the battery output 6 interferences, the highest power of which achieved -78.25 dBm; the linear power source output 6 interferences, the highest power of which achieved -67.11 dBm; the switching power source output 15 interferences, the highest power of which achieved -35.83 dBm. The overall SNRs of the battery at the 18 driving frequencies were higher than that of the linear power source, and the overall SNRs of the linear power source were higher than that of the switching power source. Conclusion The power mode of battery is the best for the hardware of EIT among the three power sources.