Design of SpO_2 Measuring and Analysis System Based on LabVIEW
- VernacularTitle:基于LabVIEW的脉搏血氧饱和度测量与分析系统设计
- Author:
Shilin HE
;
Guangfei CHEN
;
Qi LI
;
Jun YING
- Publication Type:Journal Article
- Keywords:
SpO2;
LabVIEW;
pulse signal;
MSP430
- From:
Chinese Medical Equipment Journal
2003;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To design and develop a kind of SpO2 detection and analysis system based on LabVIEW. Methods Pulse signal process and SpO2 calculation were achieved by designing a control system based on TI Company's MSP430FG439 MCU and using the digital signal process and graphic controls of the LabVIEW software. The design project and the realization way were given out by discussing the software of LabVIEW. Results The system could acquire, display and process pulse signal and calculate SpO2. Conclusion The system carries out a new way to design medical monitor equipment. Furthermore, it provides a shortcut for special analysis of the SpO2 and pulse signal in clinic.