Dynamic Air Sterilization in Newborn Hospital Ward by Laoken KDSJ-Y Multi-purpose Air Disinfecting Machine
- VernacularTitle:新生儿病房应用老肯KDSJ-Y多功能空气消毒机的效果观察
- Author:
Lixin SUN
;
Cuiqing LIU
- Publication Type:Journal Article
- Keywords:
Disinfecting machine;
Dynamic disinfection;
Newborn hospital ward;
Air sterilization
- From:
Chinese Journal of Nosocomiology
2009;0(17):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE To discuss the disinfection effect of Laoken KDSJ-Y multi-purpose air disinfecting machine using in the newborn hospital ward for air sterilization.METHODS Standard sample were carried on separately before starting,the starting 60 minutes later,the engine off latter 60 minutes and 120 minutes,after raising and the bacteria number were calculated.RESULTS Dynamic disinfection in room temperature 30-31℃,humidity 68-72%.Room population 10.7,and does not limit in the undertake transfer of personnel situation,starting disinfects 60 minutes to be able to achieveⅡ The kind of disinfection ambient air contains the fungus quantity standard;After before starting disinfection,and engine off,colony quantity continues to increase,cannot achieve toⅡ The kind of disinfection ambient air including the fungus quantity request after 120 minutes,must starting disinfect once more.CONCLUSIONS Laoken KDSJ-Y multi-purpose air disinfecting machine under some person of conditions the dynamic disinfection to be possible in the control air the bacterial content,conforms to the newborn hospital ward 2 kind of ambient air disinfection sanitation standard,suits the promotion.