Performance Evaluation of BC-5500 Hematology Analyzer
- VernacularTitle:迈瑞BC-5500全自动血细胞分析仪性能的初步评价
- Author:
Jiaxin YUE
;
Yating LAN
;
Han XU
;
Yu GAO
;
Hong ZHU
;
Yulong CONG
- Publication Type:Journal Article
- Keywords:
hematology analyzer;
precision;
reproducibility;
carryover;
linearity
- From:
Chinese Medical Equipment Journal
1993;0(06):-
- CountryChina
- Language:Chinese
-
Abstract:
0.99). DC: reproducibility was good for neutrophils, lymphocytes monocytes, eosinophils and basophils. Comparison of the results by instrument with manual for normal samples in morphology, the correlation was better for neutrophils, lymphocytes and eosinophils (r:0.968~0.983) ,good for monocytes(r=0.917), not good for basophils(r=0.659);The WBC scattergram would change and alarm flags would display when there are neutrophilic stab granulocytes, abnormal or atypical lymphocytes and immature cells in higher proportion.Conclusions The analyzer can be used to test blood cell parameters accurately and reliably. Its main performance indices accorded with the experimental requirements; The results were credible. It is necessary to check with microscopy for DC before reported when it were doubtted.