Detection of bacterial contamination with BacT/ALERT 3D in cord blood bank
- VernacularTitle:全自动微生物监测系统在脐血细菌检测中的应用
- Author:
Liwei ZHU
;
Xuemei YANG
;
Xiaoqin XU
- Publication Type:Journal Article
- Keywords:
Bact/ALERT 3D;
Umbilical cord blood;
Microbial detection
- From:
Chinese Journal of Blood Transfusion
2002;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To define a suitable and adequate sample volume for the detection of bacterial contamination in cord blood (CB) via BacT/ALERT system. Methods A total of 1 376 cord blood units were routinely screened for aerobic and anaerobic microorganisms by culturing the erythrocyte fraction(EF) or plasma fraction(PF) with 20 ml and 10 ml sample volume. The microbial contamination was detected by BacT/ALERT 3D system, and the results of this method were compared with the improved Martin/thiogly collate broth(22 ℃) and thioglycollate broth(35 ℃).Results The rates of positive BacT/ALERT results were 1.16%(10 ml) and 1.24%(20 ml) for PF, and 3.20%(20 ml) for EF. The positive rate of Martin/thiogly collate broth(22 ℃/35 ℃) was 1.00% for EF. Conclusion Using BacT/ALERT system for microbial detection in cord blood bank can enhance the security of umbilical cord blood transplantation.