Accurate Measurement of Automated Sphygmomanometer
- VernacularTitle:自动血压计的准确性评测
- Author:
Feng PAN
;
Yi WAN
;
Ying LIANG
;
Zhe YANG
;
Yongyong XU
- Publication Type:Journal Article
- Keywords:
automatic sphygmomanometer;
blood pressure measurement;
evaluation;
protocol
- From:
Chinese Medical Equipment Journal
1993;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To show the international protocols for blood pressure monitoring based on a real example. Methods The assessment process of international protocol that can be released by Working Group on Blood Pressure Monitoring of European Society of Hypertension was evaluated. Results 33 participants were selected, which all indexes in evaluation stage one and stage two of the indicators were detected through. The 95% consistency interval in difference between tested device and reference monitor was 10.65~-12.67 mmHg for systolic BP and 13.68~-14.03 mmHg for diastolic BP, and there were 7.1% (7/99) and 6.1% (6/99) of valid points out of the 95% consistency interval. Conclusion The measured automatic blood pressure in the normal environment, measuring accuracy and the standard with the control of mercury -type sphygmomanometer is coincident, so it can be recommended for home application.