Development of Developer with Fast Determining Stopped Spot of Electrolyte Electrode in Wartime
- VernacularTitle:战时快速确定电解质电极受堵部位显色剂的研制
- Author:
Yifeng ZHOU
;
Zhujun ZHOU
- Publication Type:Journal Article
- Keywords:
electrolyte;
site by plugging;
developer
- From:
Chinese Medical Equipment Journal
1989;0(02):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop developer with fast determining stopped spot of electrolyte electrode in wartime.Methods According to certain proportion configuration,the developer was prepared by using relative reagent and distilled water.Through injector to take 1ml developer and curve syringe needle,then straight insertion the type needle of electrolysis meter and pour into the developer to fill suction tube.Results The reagent can be seen various electrodes change and fast display the electrode breakdown anywhere.Conclusion The color developing reagent is low-cost,efficient and convenient to use.