Comparison Between Varian Localization System and Tosplane Localization System
- VernacularTitle:Varian定位系统与Topslane定位系统的比较
- Author:
Dequan YU
;
Hong GAO
;
Qiuju SHAO
;
Xi ZHANG
- Publication Type:Journal Article
- Keywords:
Varian localization method;
Topslane localization method;
precision error
- From:
Chinese Medical Equipment Journal
1989;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate precision error between Varian auri-point localization system and Tosplane digital-bed localization system. Methods Varian Eclipse localization method and Topslane localization method were used separately to locate one target area for one same patient. Then deviations on the tumor center coordinate were measured under the simu- lation localization machine. Results Under two localization methods, the length error: 1.7mm ?0.3mm; the width error: 0.5mm?0.4mm; the altitude error: 1.2mm?0.4mm. Conclusion The curative effect of radiotherapy would be improved by applying the Topslane localization bed to the Varian Eclipse plan system.