Simulation Study on Influence of PALL and Tyco Breathing Filter to Ppeak and Pmean
- VernacularTitle:PALL和Tyco呼吸过滤器对Ppeak和Pmean影响的模拟研究
- Author:
Xiaohai WANG
;
Dingqiu CAI
;
Bo XU
;
Lijun WANG
- Publication Type:Journal Article
- Keywords:
breathing filter;
airway resistance;
simulation study
- From:
Chinese Medical Equipment Journal
1989;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To learn the influence of PALL BB25S and Tyco Hygroboy breathing filter (BF) to Ppeak and Pmean. Methods The changes of Ppeak and Pmean were detected by using PALL BB25S and tyco Hygroboy in simulation ventila- tion. Results (1)When BFs (PALL or tyco, adult or child) were applied and Vt from 100ml/T to 800ml/T, the effect of breathing filter can be no significant changed, but Pmean was raised and Ppeak no change. (2)To compare the tyco BFs, the Pmean of child BF and adult BF, Pmean was significant increased. Conclusion Both of PALL BF and Tyco BF only significantly increase Pmean.